Examination of Effectiveness of Henniker Plasma Cleaner using a Holey Carbon Film by TEM

This work shows the effectiveness of the plasma treatment capability of the Henniker Plasma cleaner at different times and power levels.

Examination of Effectiveness of Henniker Plasma Cleaner using a Holey Carbon Film by TEM

Plasma Cleaning Method

Henniker HPT-100 Plasma Cleaner with Jeol TEM sample holder adapter.Continuously variable power output from 0-100W and dual digital MFC gas mixing. Samples were introduced via the front feedthrough adapter and cleaned with Ar:O2 95:5 mixture at low power (20W and 50W) in 30sec time steps. Results for accumulated treatment time are presented.

henniker plasma models page TEM cleaner

Fig 1. Henniker Plasma with TEM Adapter.

Preparation and Examination Methods

A holey carbon-coated Cu TEM grid.

The samples were examined in the JEM 2800 (Scanning) Transmission Electron Microscope using the following instrumental conditions: Voltage (kV) 200; C2 aperture (um) 70 and 40;

Bright-field imaging mode using CCD High magnification lattice resolution imaging mode using CCD

Examination of Effectiveness of Henniker Plasma Cleaner using a Holey Carbon Film by TEM - Analysis region henniker tem app noteFig 2&3. Analysis Region 50μm & 2μm

Cleaning progression for selected thinned analysis region showing gradual loss of carbon from the region followed by gradual loss from thicker wall regions.

Examination of Effectiveness of Henniker Plasma Cleaner using a Holey Carbon Film by TEM - Image TEM 3 rgb

Examination of Effectiveness of Henniker Plasma Cleaner using a Holey Carbon Film by TEM - Image TEM 2 rgb

Conclusion

This work shows the effectiveness of the plasma treatment capability of the Henniker Plasma cleaner at different times and power levels. The images taken from an identical location using a finder holey carbon grid showed a gradual loss of carbon from a small thinned area of carbon. First, the thinned carbon was removed (or eaten away). Then the thick walls were observed to lose carbon from the edges.

 Dr Dogan Ozkaya (Johnson Matthey Technology Centre);

“In conclusion, the Henniker HPT plasma cleaner was observed to be suitable for TEM preparation purposes because it offers amazing control over how much carbon is removed from samples. It is suitable for catalyst samples on carbon for hydrocarbon contamination cleaning in the TEM.”

Acknowledgements

We gratefully acknowledge the work of Dr Dogan Ozkaya and the talented Microscopy group members of Johnson Matthey Technology Centre who provided the detailed analyses presented in this note.

Download the application note here

REQUEST A QUOTATION OR FURTHER INFORMATION

See what our customers have to say

  • "The technical team at Henniker are very knowledgeable and supportive and always approachable. I have found it a pleasure to work with them."

    Simon Baxter - BAE Systems, MAI

    BAE
  • "Henniker guided us to choose the most suitable plasma unit for our application, ensured an accelerated delivery time & guided us through the very easy setup. We obtained quality results with their unit within minutes of setup & consistent results thereafter. The support they have provided has been rapid and helpful."

    Dr Ravi Desai - Making Lab, Francis Crick Institute

    Francis Crick Institute
  • "Henniker provided our team with excellent service during the course of our work together, the plasma cleaner arrived quickly and was installed with ease, giving us visible results from the outset and confirming that we made the right decision in choosing a local UK manufacturer."

    Dr Panagiotis Manesiotis BSc MRSC - Queen’s University Belfast

    Queens Uni Belfast
  • "Our customers and operations demand reliability at every level and were a key factor in our decision to choose a UK based manufacturer of plasma treatment equipment."

    Tom Doak - Trak Microwave

    Queens Uni Belfast
  • "Henniker’s after sales support is first class. They have always been extremely responsive if we have ever had need to call on them."

    Steve Rackham - Teledyne

    Teledyne
  • "Henniker’s plasma systems have delivered tangible benefits to us right from day one. The team there are very easy to work with."

    Ian Bruce - Coopervision

    Coopervision
  • "Henniker really stood out, both in their product range and technical knowledge. They are a great company to work with."

    Karthik Nair - University of Bradford

    University of Bradford
  • "We are very impressed with the ease of use and reliability of our plasma unit and were producing results within minutes of setting it up."

    Dr Neil Wilson - Warwick University

    Warwick University
  • "Our collaborative work with the team at Henniker was a very positive experience and one that we look forward to developing further."

    Ewen Kellar - TWI

    TWI
  • "Henniker provided a tailored product to match our exact requirements. They are a pleasure to work with."

    Dr Will Shu - Heriot Watt University

    Heriot Watt University

  • Huf
  • Kingspan
  • BOC
  • Maclaren
  • Cambridge Uni
  • Warwick Uni
  • Morgan Ceramics
  • QMUL
  • University Bradford
  • BAE
  • Imperial College London
  • Heriot Watt Uni
Henniker Plasma Logo

ISO logo

 

 

Connect with us

Henniker 3 Berkeley Court, Manor Park, Runcorn, WA7 1TQ, UK

+44 (0)1925 830 771   +44 (0)1925 800 035     info@plasmatreatment.co.uk